Replacing a hearing aid receiver is one of the most satisfying DIY repairs you can do. It takes about two minutes, requires no tools, and costs a fraction of a professional repair visit. If your Signia, Rexton, or TruHearing hearing aid has gone quiet on one side or sounds distorted after the wax guard has been changed, the receiver is the most likely cause.
Signia, Rexton, and TruHearing hearing aids all use the same Sivantos platform receiver system, so the same receivers fit all three brands. We carry receivers for IX, AX, and X platform models.
Signia, Rexton, and TruHearing are all built on the same WS Audiology Sivantos platform. Their receivers are interchangeable across brands, a Signia IX receiver fits a Rexton Reach and an equivalent TruHearing model. What matters is platform, power level, and wire length, not which brand name is on the hearing aid shell.
Receivers are the most common hardware failure in RIC hearing aids, but the wax guard is the most common cause of hearing aid silence overall. Before ordering a receiver, change the wax guard. A plugged guard causes complete silence and takes thirty seconds to fix. If sound returns after changing the guard, the receiver is fine.
Signia and Rexton have released three hearing aid platforms in recent years. The receiver connector changed between the X platform and the AX/IX platform:
AX and IX receivers share the same connector and are compatible with each other. X platform receivers use a different connector and are NOT compatible with AX or IX hearing aids. Confirm your platform before ordering.
Signia receivers come in four power levels, stamped on the wire near the connection point:
Match the letter on your current receiver exactly when ordering a replacement.
Wire length is a number (0, 1, 2, or 3) printed on the receiver wire:
The letter and number are usually printed together on the wire — for example, M2 means medium power, wire length 2.
Horizon by HearUSA and some Miracle-Ear RIC models also use Signia platform technology and may use compatible receivers. Contact us with your specific model name, and we can confirm compatibility before you order.
Red receiver wires and tips go in the right ear. Blue goes on the left. This is universal across all hearing aid brands.
Receivers don’t like moisture or wax, the two most common causes of receiver failure. If you are replacing receivers regularly, a nightly dehumidifier significantly extends their lifespan. Not sure if the receiver is the problem? Check the full troubleshooting guide first. For a complete explanation of how receivers work, see our receiver guide.
Not sure which receiver fits your Signia, Rexton, or TruHearing hearing aids? Contact Hears Hearing & Hearables with your model name and current receiver markings, and we will confirm before you order.